Invia Liberty clinician support
Please choose topic
Change pressure level
1. Enter Administrative mode, press and hold simultaneously.
2. Press selection buttons up or down to set pressure level from preset menu. For pressure level outside preset values, refer to the instructions for use - chapter "Change pressure setup mode".
3. Press "OK" to confirm and enter main display.
Change canister
1. Press “Standby” for 3 seconds.
2. Clamp canister and dressing tubing.
3. Release and remove canister (if necessary, replace canister tubing).
4. Click new Invia Liberty canister into place.
5. Unclamp canister and dressing tubing
6. Press “On” to resume therapy.
7. Seal used canister with a cap.
8. Dispose used canister in accordance with local procedures.
Choose settings: air leakage mode, language, etc.
1. Administrative mode
1. Enter Administrative mode, press and hold simultaneously.
2. Choose settings
1. Enter settings menu in Administrative mode, press and hold simultaneously.
2. Use selection buttons up or down to navigate.
3. Press "OK" to confirm setting.
Change constant/ intermittent therapy mode
1. Enter Administrative mode, press and hold simultaneously.
2. Press “Standby” for 3 seconds to put pump in standby mode.
3. Press “Change to Intermittent”. Default times for intermittent mode are 5 minutes on and 2 minutes off.
4. Press “On” to begin therapy. To change ON and OFF time in intermittent mode, refer to the instructions for use – Chapter 'Change settings'.